Publication | Open Access
Intercomparison Study of the Size-Dependent Counting Efficiency of 26 Condensation Particle Counters
166
Citations
10
References
1997
Year
Size-dependent Counting EfficiencyHigh Temperature AerosolEngineeringAerosol TransportAerosol FormationMeasurementCalibrationAerosol SamplingAir QualityParticle CountersCondensation Particle CountersEducationCalibration WorkshopParticle TechnologyInstrumentationAir PollutionIntercomparison StudyOptical Particle Sizing
ABSTRACT Particle detection efficiency curves for 26 condensation particle counters were determined during a calibration workshop in preparation for the Aerosol Characterization Experiment 1 (ACE1). Three different types of commercially available particle counters, the ultrafine condensation particle counter (TSI-3025) and the condensation particle counters (TSI-3010 and TSI-3760 or TS1-7610) were investigated at default temperature and flow settings as well as for other flow rates and temperature differences between the saturator and the condenser. Furthermore, the pulse-height-analysis ultrafine condensation particle counter and a TSI-3010 modified to achieve a higher temperature difference were calibrated. In this study, the large number of particle counters investigated provided the opportunity to obtain a more statistically significant picture of the performance of different particle counters for different operating conditions.
| Year | Citations | |
|---|---|---|
Page 1
Page 1