MedicineOrthopaedic SurgeryOrthopedic SurgeryBone GraftingJoint Replacement SurgeryOrthopedicsBone Graft MaterialsBone SubstitutesCalcium Phosphate BiomaterialsHip ArthroplastyImplant DesignJoint Replacement MaterialsFemoral Component DesignFemoral Stem DesignHydroxyapatiteHydroxyapatite BiomaterialsHydroxyapatite CoatingsPorous Hydroxyapatite
24 patients underwent bilateral hip replacement, using the SROM modular system (Depuy Johnson & Johnson, MA, USA) with the porous-coated sleeve in one hip and an HA-coated sleeve in the other. 20 patients were followed for at least 4 years with clinical and radiographic evaluations. We found no difference between the two sides as regards the time of disappearance of thigh pain and hip scores. Radiographs showed a buttress sign in 18 HA-coated and 15 porous coated stems. There were no differences in the serial bone-remodeling pattern around the proximal sleeve and measurement of wear volume between the two groups. In the short- to medium-term, our findings support the view that the hydroxyapatite-coated femoral stem has no advantages over those with porous coating.
13