Implementing a newly proposed Monte Carlo based small field dosimetry formalism for a comprehensive set of diode detectors
102
Citations
31
References
2011
Year
A number of general conclusions are evident: (1) small field OR(det) are very sensitive to the simulated source parameters, and therefore, rigorous Monte Carlo linac model commissioning, with respect to measurement, must be pursued prior to use, (2) backscattered dose to the monitor chamber should be included in simulated OR(det) calculations, (3) the corrections required for diode detectors are design dependent and therefore detailed detector modeling is required, and (4) the reported detector specific correction factors may be applied to experimental small field OR(det) consistent with those presented here.