Proceedings of the Institution of Mechanical Engineers Part H Journal of Engineering in Medicine · 2008 · 16 citations · 28 references
MedicineOrthopaedic SurgeryOrthopedic SurgeryBone GraftingBone MetabolismOrthopedicsBone BiomechanicsBone HealthBone MechanicsImplant DesignOsteoporosisVertebral FracturesAxial CompressionFemoral Stem DesignInitial Mechanical StabilityMechanical Stability AssessmentOrthopedic Implant DesignOsteoporotic FracturesTotal Hip Arthroplasty BiomechanicsWalking
The increased use of uncemented stems for hip arthroplasty and of fast-recovery protocols in elderly patients make initial stem stability and resistance to fracture critical factors in osteoporotic bone. In this paper, the subsidence and failure of two uncemented stem designs (M/L Taper and VerSys Fullcoat, Zimmer, Inc, Warsaw, Indiana, USA) in osteoporotic and non-osteoporotic cadaveric femora were compared under simulated walking conditions (axial compression and external rotation). Osteoporotic femora implanted with either stem design failed significantly more frequently than did non-osteoporotic femora. Femora implanted with the M/L stems (seven of ten by 1000 cycles) fractured earlier than did femora implanted with the Fullcoat stem (one of ten by 1000 cycles). The use of early weight-bearing protocols with uncemented stem designs in osteoporotic bone should be approached with caution.
28
Structural and cellular assessment of bone quality of proximal femur
Lawrence D. Dorr, Marie–Claude Faugere, Audley M. Mackel et al. · Bone · 1993 · 908 citations
Sudden Death During Primary Hip Arthroplasty
Javad Parvizi, Allan D. Holiday, Mark H. Ereth et al. · Clinical Orthopaedics and Related Research · 1999 · 271 citations