Magnetic Resonance in Medicine · 2008 · 41 citations · 17 references
MedicineBiomedical EngineeringRadiologyBiomedical ImagingDiagnostic ImagingDiagnostic RadiologyMagnetic Resonance ImagingMagnetic Resonance Imaging PhysicsMagnetic Resonance Signal FormationMedical ImagingMedical Imaging PhysicsCapacitanceCoil DesignConformal MappingField HomogeneityHigh-field Magnetic Resonance ImagingMagnetic Field HomogeneityRadiofrequency Coil DesignRadiofrequency Coils
An asymmetric quadrature birdcage body coil for hyperpolarized (HP) (3)He lung imaging at 1.5T is presented. The coil is designed to rest on top of the patient support and be used as a temporary insert in a clinical system. A two-part construction facilitates patient access and the asymmetric design makes maximal use of available bore space to ensure comfort. Highly homogeneous, circularly polarized RF magnetic fields are produced at 48.5 MHz using a conformal mapping method for the geometrical design, combined with an algebraic method to calculate the individual capacitance values on the birdcage coil's ladder network. Efficiency and isolation from the system's proton body coil are ensured by an integrated RF screen. The design methodology is readily applicable to other field strengths or nuclei. Improvements over existing (3)He coils were found in terms of sensitivity and transmit field homogeneity, an important feature in HP MRI.
17
Gradient-Echo Imaging Considerations for Hyperpolarized129Xe MR
Lei Zhao, Robert V. Mulkern, Ching‐Hua Tseng et al. · Journal of Magnetic Resonance Series B · 1996 · 200 citations · Full text
Resonant Modes of the Birdcage Coil
Mark C. Leifer · Journal of Magnetic Resonance · 1997 · 110 citations