Measurements of coronary velocity and reactive hyperemia in the coronary circulation of humans.

280

Citations

19

References

1981

Year

Abstract

An acceptable method for measuring phasic coronary velocity and reactive hyperemia in humans has not been available. We have developed a doppler probe which can be coupled to surface coronary vessel* at the time of cardiac surgery with a small suction cup. Phasic coronary velocity can be measured with a signal to noise ratio that exceeds 20:1. Animal studies have shown that the probe does not alter myocardial perfusion or cause tissue damage. In addition, changes in mean coronary velocity are closely related (r 0.97) to changes in coronary flow over a wide range (15-400 ml/min). The characteristics of reactive hyperemia in the coronary circulation of dogs determined with the doppler system are similar to those obtained simultaneously with an electromagnetic flow meter. Transient occlusions of branch coronary vessels in patients with normal coronary arteries are not associated with significant changes in heart rate, left atrial, or mean arterial pressure. The characteristics of reactive hyperemia in normal vessels of 13 patients were as follows: although reactive hyperemia responses were demonstrable following 1 to 2-aecond coronary occlusions, maximal responses usually occurred with 20-second coronary occlusions; following 20 seconds of coronary occlusion, the ratio of peak to resting velocity was 5.8 0.6 (mean SE); the ratio of repayment to debt area was 3.1 0.2, and the duration of the reactive hyperemia response was 20.8 0.3 seconds. These studies provide the first quantitative measurements of coronary reactive hyperemia in humans. Circ Res 49: [877][878][879][880][881][882][883][884][885][886][887][888][889][890][891] 1981 Although mean flow to the left ventricle can be assessed in patients with several techniques (Berne and Rubio, 1979;

References

YearCitations

Page 1