Magnetic Resonance Imaging · 1997 · 50 citations · 10 references
MedicineNeurosciencePhysiologyBiomedical ImagingDiagnostic ImagingDiagnostic RadiologyBrain ImagingCerebral Blood FlowMagnetic Resonance ImagingMagnetic Resonance Pulse SequencesNeuroimaging PhysicsPerfusion ImagingCerebral Blood Flow AlterationsCerebral Blood Flow QuantificationIn Vivo ImagingInversion Recovery PulsesInversion Recovery SequencesMagnetic Resonance Flow ImagingPerfusion Imaging SequencesPulse Sequence Design
A new pulse sequence for estimating cerebral blood flow called UNFAIR, which uses a combination of sequential hyperbolic secant preparatory pulses, is introduced. This sequence is based on the same generalized conditions as previously introduced inversion recovery techniques except that the spins in the image slice of interest always have +z magnetization and the in-flowing spins are alternately inverted and uninverted. CBF-weighted images of rat brain under conditions of normocpnia and hypercapnia are presented and demonstrate the expected CBF response. A model describing the signal response to this pulse sequence is also presented and compared with in-vivo data acquired from gray and white matter.
10
John A. Detre, John S. Leigh, Donald S. Williams et al. · Magnetic Resonance in Medicine · 1992 · 1.6K citations
Medicine, Neuroscience, Radiology +16
Mr perfusion studies with t1‐weighted echo planar imaging
Kenneth K. Kwong, David A. Chesler, Robert M. Weisskoff et al. · Magnetic Resonance in Medicine · 1995 · 476 citations
Medicine, Neurology, Neuroscience +17
Selective population inversion in NMR
Michael S. Silver, R. I. Joseph, C.-N. Chen et al. · Nature · 1984 · 209 citations