Publication | Closed Access
The Influence of Central Corneal Thickness and Corneal Curvature Radius on The Intraocular Pressure as Measured By Different Tonometers: Noncontact and Goldmann Applanation Tonometers
56
Citations
16
References
2008
Year
NCT can be more affected by CCT than GAT. CCR could influence the measurements of GAT, but not those by NCT. CCT and CCR can therefore influence the discordance of IOP readings taken with NCT and GAT.