Concepedia

Publication | Open Access

Effects of a Circulating-water Garment and Forced-air Warming on Body Heat Content and Core Temperature

119

Citations

28

References

2004

Year

Abstract

The circulating-water system transferred more heat than forced air, with the difference resulting largely from posterior heating. Circulating water rewarmed patients 0.4 degrees C/h faster than forced air. A substantial peripheral-to-core tissue temperature gradient with each device indicated that peripheral tissues insulated the core, thus slowing heat transfer.