Journal of Liquid Chromatography & Related Technologies · 1997 · 19 citations · 21 references
PharmacyChemistryPharmacologyAnalytical ChemistrySeparation ProcessesSeparation ScienceChromatographic Analysis Of PharmaceuticalsChromatographic SeparationChromatographyLiquid ChromatographyMobile Phase OptimizationMobile Phase SelectionAnalysis Time OptimizationChromatographic Drug AnalysisColumn Temperature EffectsOptimum TemperatureReversed-phase ChromatographyReversed-phase Liquid ChromatographySeparation Efficiency OptimizationTemperature Optimum
Abstract A chemometric methodology was used to study column efficiency and the separation of 10 benzodiazepines in reversed phase liquid chromatography. New simple mathematical models and the organic modifier (OM) organization of ACN in the water, explained differences on column efficiency observed when ACN is chosen instead of CH3OH. A new response function, which takes into account the separation quality and the analysis time, was proposed for the separation optimization. The result, a mobile phase ACN/water (60/40)(V/V), with a flow rate = 1.00 mL/min and a column temperature = 47°C were optimum values for a rapid chromatographic separation.
21
Joseph L. Glajch, J. J. Kirkland, Karen M. Squire et al. · Journal of Chromatography A · 1980 · 490 citations