IEEE Transactions on Nuclear Science · 1977 · 31 citations · 3 references
EngineeringNuclear EngineeringPhysicsNuclear PhysicsRadiation DetectionRadiation PhysicsDetector TechnologyDislocation MechanicsEnergy ResolutionGamma-ray DetectionGamma-ray DetectorsRadiation DetectorsDetector DesignDetector Energy ResolutionDislocation DensityDislocation PhysicsPlanar DetectorsRadiation DefectsRadiation Energy ResolutionUniformity Analysis
The energy resolution and response uniformity of HgI2 detectors to nuclear radiation depends strongly on dislocation density and distribution in the HgI2 crystal bulk. Several large planar detectors fabricated from the interiors of large single crystals verified this dependence. Scanning with collimated low-energy (6 keV) and medium-energy (60 keV) gamma-rays confirmed the association of poor resolution and counting efficiency with regions of high dislocation density and crystal imperfection.
3
Properties of Vapour Phase Grown Mercuric Iodide Single Crystal Detectors
J.P. Ponpon, R. Stuck, P. Siffert et al. · IEEE Transactions on Nuclear Science · 1975 · 78 citations