Comparison of film dosimetry techniques used for quality assurance of intensity modulated radiation therapy

17

Citations

26

References

2010

Year

Abstract

The EDR2 film WF and GafChromic EBT film achieved an average accuracy level of 1.5% against an ionization chamber. These two techniques agreed with the MC prediction in 2%/2mm criteria evaluated by the gamma index, whereas EDR2 without filters achieved an accuracy level of 3%/3 mm with the decision criteria of agreement greater than 95% of the tested region. The overall results will provide a useful quantitative reference for IMRT verifications.