Rugged and Accurate Quantitation of Colchicine in Human Plasma to Support Colchicine Poisoning Monitoring by Using Turbulent-Flow LC–MS/MS Analysis

E. Bourgogne, Marion Soichot, Colette Latour, Olivier Laprévôte

Bioanalysis · 2013 · 14 citations · 21 references

Concepts

Abstract

BACKGROUND: Colchicine is a common drug used in inflammatory diseases. The narrow therapeutic index requires fast and reliable techniques for its quantitation. An online, automated sample preparation using TurboFlow™ technology combined with triple-stage quadrupole MS detection was applied to identify colchicine in human plasma and follow intoxications. METHODOLOGY: Plasma samples (200 µl) were mixed with deuterated colchicine and protein precipitation ZnSO4 solutions. After centrifugation, supernatants were extracted onto a Cyclone P TurboFlow column and eluted onto a narrowbore Hypersil™ GOLD column with a methanol/water gradient. Analytes were monitored in SRM mode (positive electrospray). RESULTS: Total run time was 9.5 min. Calibration curves ranged from 0.342 to 17.1 ng/ml, with significant linearity (R(2) >0.99). Inter- and intra-assay precisions were <16.8% and accuracy was 84.4-110%. CONCLUSION: This method is suitable for monitoring intoxication in patients undergoing chronic treatment and is routinely applied to toxicological samples.

References

21