Physics in Medicine and Biology · 1999 · 75 citations · 26 references
EngineeringNuclear EngineeringNuclear InstrumentationRadiation DosimetryRadiation MeasurementAbsorbed DoseDose CalibrationDosimetry InstrumentationIonization Chamber DosimetryIonization ChambersRadiation Dose MeasurementAbsorbed Dose-to-water DeterminationExposure MeasurementHeavy Ion RadiationIon Beam DosimetryIonizing Radiation DosimetryIonizing Radiation Exposure AssessmentRadiation Dose AssessmentThimble Ionization Chambers
The method to measure absorbed dose to water in a field of carbon ions as applied for the heavy ion therapy project at the Heavy Ion Research Laboratory in Darmstadt (GSI), Germany, is described in detail. Thimble ionization chambers with a water absorbed calibration factor are applied. The dose obtained with this method was compared with that obtained at the heavy ion therapy facility HIMAC at the National Institute of Radiological Sciences in Chiba, Japan, using the Japanese code of practice. The agreement found was better than 1%. The combined uncertainty of the determination of absorbed dose to water was estimated to amount to 5%.
26
A measurement of → (1975 ⩽ ⩽ 2250)
D. Bisello, S. Limentani, M. Nigro et al. · Nuclear Physics B · 1983 · 112 citations