Medical Physics · 1980 · 24 citations · 0 references
EngineeringNuclear EngineeringPhysicsRadiation DosimetryRadiation PhysicsRadiation ShieldingAccelerator PhysicsAccelerator ShieldingCharged Particle ShieldingElectron DosimetryIonization ChambersShielding For Particle AcceleratorsAccelerator Radiation Shielding DesignElectron ShieldingIonization Chamber DesignPatient Dose ReductionShielding For Electron AcceleratorsShielding For Linear AcceleratorsShielding Performance EvaluationShielding Thickness Calculations
The relative dose reduction by Lipowitz metal of 6 to 20 MeV electrons from a Varian Associates Clinac-20 linear accelerator has been measured using a parallel plate thin wall ionization chamber. Metal thickness required for a 5% attenuation level for a 10 X 10 cm2 field are as follows: 6 MeV-2.3 mm, 9 MeV-4.4 mm, 12 MeV-8.5 mm, 16 MeV--18.0 mm, 20 MeV-25.0 mm.