Improved apparatus for the NMR measurement of self-diffusion coefficients using pulsed field gradients

D.S. Webster, K. H. Marsden

Review of Scientific Instruments · 1974 · 57 citations · 10 references

Concepts

Abstract

The characteristics of some quadrupole coils are examined to determine their suitability for NMR pulsed field gradient diffusion measurements. They are found to be capable of acceptable field gradient uniformity, low ratio of inductance to field gradient strength, low dc resistance and their windings are as far as possible from adjacent pole faces, thus minimising eddy current formation. An improved current pulser using SCR's is also discussed.

References

10