Journal of Interventional Cardiac Electrophysiology · 2011 · 23 citations · 10 references
Clinically relevant particles, both visible and sub-visible, with the potential for causing embolic complications, are generated by the BRK-1 needle without stylet. Use of a stylet in the BRK-1 needle, or the BRK-1XS needle with or without stylet, appears to reduce the size and amount of particles created. The NRG C1 needle appears to eliminate visible particles and is comparable to the BRK-1 with stylet and the BRK-1XS with or without stylet in generation of sub-visible particles. Important steps can be taken to minimize the creation of particles during the advancement of the BRK-1 through the transseptal sheath and dilator.