Medical Physics · 2003 · 33 citations · 9 references
MedicineBiomedical EngineeringRadiologyDiagnostic ImagingDiagnostic RadiologyRadiation DetectionBreast ImagingMammography PhysicsMedical ImagingRadiographic ImagingAnti-scatter GridsContrast-to-noise Ratio EstimationFull-field Digital MammographyMammographyMammography ImagingPatient Dose ReductionScatter In MammographySignal-to-noise Ratio
Effects of x-ray scattering on full-field digital mammography are analyzed with the scattering model of Seibert and Boone [Med. Phys. 15, 567-575 (1988)]. A new method is introduced for the estimation of model parameters from measurements. It is shown that with breasts thinner than a certain threshold, removing the anti-scatter grid leads to an improved contrast-to-noise ratio with a smaller patient dose. A fast approximate algorithm is presented for determining the scattered field in a gridless digital mammogram.