Screw stimulation monitoring is a valuable and efficacious adjunct to lumbar pedicle screw instrumentation. A stimulation threshold greater than 15 mA reliably indicates adequate screw position. A stimulation threshold between 10 and 15 mA was generally associated with adequate screw position, although exploration of the pedicle is recommended. A stimulation threshold less than 10 mA was associated with a significant cortical perforation in most instances.