Proceedings of the Institution of Mechanical Engineers Part H Journal of Engineering in Medicine · 2006 · 16 citations · 15 references
MedicineOrthopaedic SurgeryOrthopedic SurgeryJoint Replacement SurgeryOrthopedic BiomechanicsOrthopedicsArthroplasty BiomechanicsBearing Surface MaterialsHip ArthroplastyHip BiomechanicsImplant BiomechanicsJoint Replacement MaterialsDry Friction BehaviorHip Contact MechanicsHip Joint StabilityImplant Wear DebrisJoint Contact MechanicsMetal Wear DebrisProsthesis Wear MechanismsWear Mechanics
The contact mechanics of two metal-on-metal (MOM) total hip prostheses was studied by means of the finite element method (FEM). The purpose of the work was to compare two total hip replacements (Durom and Metasul) with regard to the amount of wear debris released. Wear on the bearing surfaces was evaluated following Reye hypotheses from the pressure distribution, computed by means of three-dimensional FEM models; an approximate analytical model based on Hertz contact theory has also been developed and discussed. The results show that in the dry friction condition the Durom joint releases almost twice as much wear volume as produced by the Metasul joint. Therefore, while Durom implants can improve hip stability by increasing the prosthetic impingement-free range of motion (PIF-ROM), Metasul prostheses can be a valuable solution whenever wear represents a critical choice factor.
15
Biomechanics of large femoral heads: what they do and don't do.
Roy D. Crowninshield, William J. Maloney, Douglas H. Wentz et al. · PubMed · 2004 · 199 citations
Biomechanics of Large Femoral Heads
Roy D. Crowninshield, William J. Maloney, Douglas H. Wentz et al. · Clinical Orthopaedics and Related Research · 2004 · 192 citations