Publication | Closed Access
Arterial to endtidal carbon dioxide gradient during pediatric laparoscopic fundoplication
18
Citations
9
References
2008
Year
ETCO(2) may not accurately represent arterial values during laparoscopic fundoplication, especially in the infant when carbon dioxide insufflation is used. Consideration should be given to placing an arterial line for blood gas measurement in some patients.