Publication | Closed Access
Blood leak alarm interference by hydoxocobalamin is hemodialysis machine dependent
18
Citations
7
References
2012
Year
We alert physicians to become familiar with the properties of individual dialysis machines prior to use of hydroxocobalamin. When facing difficulties with hemodialysis after the administration of hydroxocobalamin, consider attempting with a different manufactures machine or model if available or contact the manufacturer directly.