Titanium (Ti) alloy metal has been used for permanent implant in the human body. Its high strength and hardness take place due to the β phase formation at room temperature. Ti-6%Al-6%V alloy is most popular, of which vanadium (V) content is used as the β phase stabilizer. However V can induce allergic reaction from the body. V can be substituted by other element, such as Mo and Nb. Microstructure observations for Ti-6%Al-6%Mo and Ti-6%Al-6%Nb alloys show that β phase exists as matrix having good workability at room temperature. After preheat at 1000°C, no cracking failure occurs during forging and rolling treatment. Thermal spray method is used for CaPO 4 surface treatment. A CaPO 4 layer on the alloy substrate forms after the molten CaPO 4 is hot sprayed on to the alloy substrate surface. Corrosion test results indicate that the increase in Mo or Nb content up to 6% leads to the increase in corrosion resistance.