Precision Adiabatic Gamma-Ray Calorimeter using Thermistor Thermometry

I. T. Myers, W. H. Le Blanc, D.M. Fleming

Review of Scientific Instruments · 1961 · 10 citations · 1 references

Concepts

Abstract

A high precision adiabatic gamma-ray calorimeter is described. For most gamma-ray sources, the calorimeter walls are thick enough to absorb more than 99% of the gamma rays emitted by the source being measured. The standard deviation of the mean in the measurement of a half-curie Co60 source was 0.2%; that of a half-curie radium source was 0.03%.