MedicineEndocrinologyInternal MedicineBioinstrumentationDiabetologyMetabolic RegulationDiabetes ManagementDiabetes MellitusGlycemic ControlGlycemic MonitoringHypoglycemia ManagementInsulin TherapyBasal Insulin TherapyHypoglycemia In DiabetesInsulin Delivery DevicesInsulin PumpsInsulin Therapy ResponseInsulin-associated HypoglycemiaInsulin-mediated HypoglycemiaNocturnal Hypoglycemia
OBJECTIVE: To evaluate a sensor-augmented insulin pump with a low glucose suspend (LGS) feature that automatically suspends basal insulin delivery for up to 2 h in response to sensor-detected hypoglycemia. RESEARCH DESIGN AND METHODS: The LGS feature of the Paradigm Veo insulin pump (Medtronic, Inc., Northridge, CA) was tested for 3 weeks in 31 adults with type 1 diabetes. RESULTS: There were 166 episodes of LGS: 66% of daytime LGS episodes were terminated within 10 min, and 20 episodes lasted the maximum 2 h. LGS use was associated with reduced nocturnal duration ≤2.2 mmol/L in those in the highest quartile of nocturnal hypoglycemia at baseline (median 46.2 vs. 1.8 min/day, P = 0.02 [LGS-OFF vs. LGS-ON]). Median sensor glucose was 3.9 mmol/L after 2-h LGS and 8.2 mmol/L at 2 h after basal restart. CONCLUSIONS: Use of an insulin pump with LGS was associated with reduced nocturnal hypoglycemia in those at greatest risk and was well accepted by patients.