Physics in Medicine and Biology · 2004 · 75 citations · 12 references
MedicineBiomedical EngineeringRadiologyDiagnostic ImagingDiagnostic RadiologyRadiation DosimetryBreast Cancer ImagingBreast ImagingCancer DetectionComputed TomographyMammography PhysicsMedical ImagingCone-beam Computed Tomography (Cbct)Flat-panel Detector PerformanceFlat-panel Detector SystemsFlat-panel DetectorsMammographyMammography DetectorsMammography ImagingScreening Mammography
The purpose of this study was to investigate microcalcification detectability using CT mammography with a flat-panel imager. To achieve this, a computer simulation was developed to model an amorphous-silicon, CsI based flat-panel imager system using a linear cascaded model. The breast was modelled as a hemi-ellipsoid shape with composition of 50% adipose and 50% glandular tissue. Microcalcifications were modelled as small spheres having a composition of calcium carbonate. The results show that with a mean glandular dose equivalent to that typically used in two-view screening mammography, CT mammography with a flat-panel detector is capable of providing images where most microcalcifications are detectable. A receiver operating characteristic (ROC) study was conducted by five physicist observers viewing simulated CT mammography reconstructions. The results suggest that the microcalcification with its diameter equal to or greater than 0.175 mm can be detected with an average area under the ROC curve (AUC) greater than 0.95 using 0.1 or 0.2 mm pixelized detectors. The results also indicate that the optimal pixel size of the detector is around 0.2 mm for microcalcification detection, based on the trade-off between detectability of microcalcifications and the time required for data acquisition and reconstruction.
12
L.A. Feldkamp, L. C. Davis, J. W. Kress · Journal of the Optical Society of America A · 1984 · 6.2K citations
Dedicated Breast CT: Radiation Dose and Image Quality Evaluation
John M. Boone, Thomas R. Nelson, Karen K. Lindfors et al. · Radiology · 2001 · 464 citations
Anthony B. Wolbarst · Health Physics · 2004 · 299 citations
Arts, Radiology, Biomedical Imaging +11