Optics Express · 2010 · 17 citations · 13 references
EngineeringBiomedical EngineeringSurgeryBiomedical OpticsBiophotonicsOpticsBiomedical Laser ApplicationsCorneal OpticsFemtosecond Laser ParametersFemtosecond Laser PulsesLaser Cutting ParametersNonlinear RefractionNonlinear Refractive IndexNonlinear Wave RefractionPorcine ModelsRefraction
We have investigated the nonlinear refractive index of ex-vivo pig cornea by a combined approach using the standard z-scan technique on extracted corneas or corneal slices, as well as studying the deviations caused by self-focusing during femtosecond laser processing of the pig eyes. The experiments yield consistently an upper limit of 1.2 MW for the critical power of self-focusing in porcine cornea, and a value of 2 x 10(-19) m(2)/W for its nonlinear refractive index. We also demonstrate that due to this nonlinear refraction the cutting depth of typical fs-laser surgery processing in cornea may depend considerably, albeit in a well controllable way, on the laser parameters.