Publication | Open Access
Measurement accuracy improvement with PZT scanning for detection of DPC in Hi-Bi fibers
18
Citations
0
References
2002
Year
EngineeringMeasurementHi-bi FibersEducationFiber OpticsFiber-optic CommunicationPolarization-coupling IntensityDistributed PolarizationHigh-performance FiberOptical PropertiesCalibrationPiezoelectric MaterialInstrumentationPiezo-electronic TransducerFiber Optic SensingAcoustic PropagationPiezoelectricityUltrasoundFiber OpticPolarization ImagingMeasurement Accuracy ImprovementPiezoelectric NanogeneratorsTransducer PrincipleApplied PhysicsFibre Amplifier
Piezo-electronic transducer is used in combination with mechanical scanning devices to improve measurement accuracy in a distributed polarization coupling detection system. For mechanical scanning range of 150 mm with 200-nm resolution, the measurement error of polarization-coupling intensity can be improved from 38% to 2% in combination with a 20-nm resolution piezo-electronic transducer.