IEEE Transactions on Nuclear Science · 2014 · 45 citations · 10 references
Natural SciencesMaterials SciencePhysicsAnalytical ChemistryRadiation DetectionRadiation MeasurementDetector TechnologyGamma-ray DetectionRadiation DetectorsRadioactive DecayScintillation DetectorsScintillator MaterialsDecay TimePosition
Organic scintillators are often chosen as radiation detectors for their fast decay time and their low Z, while inorganic ones are used when high light yields are required. In this paper we show that a para-terphenyl based detector has a blend of properties of the two categories that can be optimal for energy and position measurements of low-energy charged particles. Using 0.1% diphenylbutadiene doped para-terphenyl samples we measured a light attenuation length λ = 4.73 ±0.06 mm, a quenching factor for α particles Q <sub xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">α</sub> = (10.7 ±0.6), and a rejection power ranging between 3 - 11% for 660 keV photons, with respect to electrons of the same energy, depending on the signal threshold. A simulation based on FLUKA properly reproduces the experimental data distributions.
10
The FLUKA code: description and benchmarking
G. Battistoni, F. Cerutti, A. Fassò et al. · AIP conference proceedings · 2007 · 1.3K citations
Richard A. Keller · Journal of the Franklin Institute · 1968 · 553 citations
Study of Organic Scintillators
Raymond Sangster, John W. Irvine · The Journal of Chemical Physics · 1956 · 234 citations