High-Speed Gas Extraction of Volatile and Semivolatile Organic Compounds from Aqueous Samples

Carrie Leonard, Huafen Liu, Stephen Brewer, Richard Sacks

Analytical Chemistry · 1998 · 12 citations · 13 references

Concepts

Abstract

A device is described for high-speed gas extraction of volatile and semivolatile organic compounds from aqueous samples. High-speed extraction is achieved by the use of elevated temperatures. Large quantities of water vapor produced at elevated temperatures are managed with a reflux condenser, which efficiently removes water vapor without sample loss by returning the condensed water along with any cocondensed organic compounds to the extraction cell. The heated extraction cell uses a 1−2-mL sample with an extraction gas flow of typically 30 mL/min. Extraction temperatures as high as 95 °C can be used while maintaining a dew point temperature of ∼5 °C for the extraction gas and sample vapor leaving the device. Extraction profiles are obtained by connecting the device directly to a flame ionization detector. Extraction profiles for benzene show that quantitative recovery can be achieved in ∼30 s at an extraction temperature of 90 °C. Large increases in recovery at higher extraction temperatures are also demonstrated for tridecane and ethyl alcohol. For analytical studies, the device is interfaced to a commercial cryofocusing inlet system for high-speed gas chromatography. If all the extraction gas is trapped and injected into the separation column, detection limits for volatile organic compounds typically are in the 50−200 parts-per-trillion range.