Concepedia

Publication | Closed Access

High-resolution imaging without iteration: a fast and robust method for breast ultrasound tomography

100

Citations

51

References

2011

Year

Abstract

Breast ultrasound tomography has the potential to improve the cost, safety, and reliability of breast cancer screening and diagnosis over the gold-standard of mammography. Vital to achieving this potential is the development of imaging algorithms to unravel the complex anatomy of the breast and its mechanical properties. The solution most commonly relied upon is time-of-flight tomography, but this exhibits low resolution due to the presence of diffraction effects. Iterative full-wave inversion methods present one solution to achieve higher resolution, but these are slow and are not guaranteed to converge to the correct solution. Presented here is HARBUT, the hybrid algorithm for robust breast ultrasound tomography, which utilizes the complementary strengths of time-of-flight and diffraction tomography resulting in a direct, fast, robust and accurate high resolution method of reconstructing the sound speed through the breast. The algorithm is shown to produce accurate reconstructions with realistic data from a complex three-dimensional simulation, with masses as small as 4 mm being clearly visible.

References

YearCitations