EngineeringNuclear EngineeringDiagnostic ImagingRadiation DetectionRadiation MeasurementDetector TechnologyRadiation Imaging DetectorsRadiographic ImagingX-ray DetectionX-ray DetectorsX-ray OpticsCompton DetectorsCompton ImagingDetective Quantum EfficiencyQuantum EfficiencyX-ray Imaging PerformanceX-ray Imaging Physics
Recently developed diagnostic techniques at LLNL require recording backlit images of extremely dense imploded plasmas using hard x-rays, and demand the detector to be sensitive to photons with energies higher than 50 keV [R. Tommasini et al., Phys. Phys. Plasmas 18, 056309 (2011); G. N. Hall et al., "AXIS: An instrument for imaging Compton radiographs using ARC on the NIF," Rev. Sci. Instrum. (these proceedings)]. To increase the sensitivity in the high energy region, we propose to use a combination of two MCPs. The first MCP is operated in a low gain regime and works as a thick photocathode, and the second MCP works as a high gain electron multiplier. We tested the concept of this dual MCP configuration and succeeded in obtaining a detective quantum efficiency of 4.5% for 59 keV x-rays, 3 times larger than with a single plate of the thickness typically used in NIF framing cameras.