Publication | Closed Access
Optimization of Contrast-Enhanced MR Angiography of the Hands with a Timing Bolus and Elliptically Reordered 3D Pulse Sequence
27
Citations
14
References
2000
Year
EngineeringBolus Arrival TimeElliptically Reordered 3DBolus AdministrationMagnetic Resonance ImagingTiming BolusSequence SettingContrast-enhanced Mr AngiographyCardiologyRadiologyHealth SciencesCardiovascular ImagingVascular ImageMedical ImagingAnesthesiologyContrast AgentMri-guided Radiation TherapyDigital Subtraction AngiographyBiomedical Imaging3D Imaging
Our objective was to optimize bolus administration and sequence setting in gadolinium-enhanced magnetic resonance (MR) angiography of the hands. Elliptically reordered three-dimensional (3D) spoiled gradient-echo sequence with non-slab-selective radio frequency excitation was optimized according to the measurements of arterial and venous time-signal curves in 21 patients. Great variations in bolus arrival time and arterio-venous transit time could be observed. In most patients high-quality arterial depiction could be obtained with minor venous contamination. Contrast-to-noise, spatial resolution, and selective arterial filling is still a challenge for 3D MR angiography of the hand but can be optimized using Gadolinium-BOPTA and a dedicated pulse sequence setting with exact bolus timing.
| Year | Citations | |
|---|---|---|
Page 1
Page 1