Physics in Medicine and Biology · 1997 · 290 citations · 29 references
A solid tissue phantom made of agar, Intralipid and black ink is described and characterized. The preparation procedure is fast and easily implemented with standard laboratory equipment. An instrumentation for time-resolved transmittance measurements was used to determine the optical properties of the phantom. The absorption and the reduced scattering coefficients are linear with the ink and Intralipid concentrations, respectively. A systematic decrease of the reduced scattering coefficient dependent on the agar content is observed, but can easily be managed. The phantom is highly homogeneous and shows good repeatability among different preparations. Moreover, agar inclusions can be easily embedded in either solid or liquid matrixes, and no artefacts are caused by the solid-solid or solid-liquid interfaces. This allows one to produce reliable and realistic inhomogeneous phantoms with known optical properties, particularly interesting for studies on optical imaging through turbid media.
29
Light scattering in lntralipid-10% in the wavelength range of 400–1100 nm
Hugo J. van Staveren, Christian J. M. Moes, Jan van Marie et al. · Applied Optics · 1991 · 1.1K citations
Spectroscopy and Imaging with Diffusing Light