Physics in Medicine and Biology · 1967 · 12 citations · 4 references
Natural SciencesNuclear MedicineRadiologyDiagnostic ImagingOpticsRadiation DetectionDetector TechnologyGamma-ray DetectionGamma-ray DetectorsGamma-ray InstrumentationImage QualityRadiation Imaging DetectorsGamma Camera SystemsGamma Detector SystemsGamma-ray Detector ResolutionGamma-ray Detector TechnologyGamma-ray ImagingImage Intensifier SystemsImaging ResolutionNuclear Medicine Detectors
A stationary gamma ray detector incorporating a light image intensifier is described. The light emitted from a 2 in. diameter, 3 mm thick, flat crystal is focused, by means of a special Perspex lens system, onto the photocathode of the image intensifier (also 2 in. diameter) with a 5% efficiency. Using a parallel-hole lead collimator 2 in. long and having 2 mm square holes and 0·007 in. thick septa, it was found that, when using a 99mTc source (140 kev), a resolution of 1 mm was obtainable, although contrast was poorer at this value, due in part to the acceptance of unwanted Compton events within the subject, collimator and crystal. Contrast at separations of the order of 2 mm and above was quite good, however.
4
Hal O. Anger · Review of Scientific Instruments · 1958 · 791 citations
SCINTILLATION CAMERA WITH MULTICHANNEL COLLIMATORS.
Hal O. Anger · PubMed · 1964 · 330 citations · Full text