Physics in Medicine and Biology · 2012 · 234 citations · 27 references
MedicineNuclear MedicineRadiologyDiagnostic ImagingDiagnostic Nuclear MedicineDiagnostic RadiologyAttenuation CorrectionImagingMedical ImagingNuclear Medicine ImagingPositron Emission TomographyPositron Emission Tomography PhysicsAttenuation Coefficient EstimationDirect MeasurementPet Attenuation CorrectionPet Image ReconstructionPet QuantificationPositron Emission Tomography Attenuation CorrectionTime-of-flight PetTime-of-flight Positron Emission Tomography
In positron emission tomography (PET), a quantitative reconstruction of the tracer distribution requires accurate attenuation correction. We consider situations where a direct measurement of the attenuation coefficient of the tissues is not available or is unreliable, and where one attempts to estimate the attenuation sinogram directly from the emission data by exploiting the consistency conditions that must be satisfied by the non-attenuated data. We show that in time-of-flight PET, the attenuation sinogram is determined by the emission data except for a constant and that its gradient can be estimated efficiently using a simple analytic algorithm. The stability of the method is illustrated numerically by means of a 2D simulation.
27
Choice Reviews Online · 2008 · 1.1K citations
Attenuation correction for a combined 3D PET/CT scanner
Paul E. Kinahan, D.W. Townsend, Thomas Beyer et al. · Medical Physics · 1998 · 933 citations
MRI-Based Attenuation Correction for PET/MRI Using Ultrashort Echo Time Sequences
Vincent Keereman, Yves Fierens, Tom Broux et al. · Journal of Nuclear Medicine · 2010 · 436 citations