Ultratrace and isotope ratios analyses of some radionuclides by ICP-MS
16
Citations
36
References
2004
Year
EngineeringRadioactive ContaminationMass InterferencesChemistrySr Isotope RatioEnvironmental ChemistryAnalytical ChemistryNuclear DecayChromatographyIsotope AnalysisAccelerator Mass SpectrometryRadioactive Waste DisposalEnvironmental RadiochemistryEnvironmental EngineeringIsotope GeochemistryMass SpectrometryRadioanalytical ChemistryStable Isotope ProbingGeochemistryAm DetectionMedicine
Abstract Extensive work is under way using High Resolution-ICP-MS for 90 Sr, 234 U, 235 U, 238 U, 239 Pu, 240 Pu, and 241 Am detection. Sample preparation procedures based on liquid-liquid extraction and ion exchange chromatography were developed. Sr, U, Pu, and Am were separated from their matrix and concentrated to improve the power of detection in the mass spectrometer. A microconcentric nebulizer with a desolvation introduction system (Ardius) is used. Instrumental limits of detection using Sr and U standard solutions are 0.01 ppt and 0.006 ppt for Sr and U, respectively. A study is presented on the mass interferences for the specified radionuclides. In the environmental samples investigated the 90 Sr/ 86 Sr isotope ratio is 6.02×10 -9 and for 240 Pu/ 239 Pu the isotope ratio is 0.17.