Review of Scientific Instruments · 1950 · 32 citations · 3 references
EngineeringPhysicsNuclear PhysicsRadiation MeasurementRadiation PhysicsNeutron DetectionProportional CountersCountersDetector ConstructionEfficiencyGas Proportional CountersIrradiation ConditionsNeutron Detector Efficiency
Boron trifluoride proportional counters of calculable efficiency for neutron beam measurements are described. It has been found possible to obtain excellent pulse-size distributions with a filling of 60 cm of purified BF3 by ``soaking'' the counters for a long time (up to three months) with BF3 before putting in the final filling. These fillings are stable. Fillings of less than 5 cm BF3 made up to 50 cm with argon have also been studied. With these fillings the pulse-size distributions are very good for both beam and total irradiation conditions. A discussion of the counter efficiencies with particular reference to end correction is given.
3
Neutron Measurements with Boron-Trifluoride Counters
S. A. Korff, W.E. Danforth · Physical Review · 1939 · 58 citations