Concepedia

Publication | Closed Access

Fixation of Periprosthetic Femur Fractures: A Biomechanical Analysis Comparing Cortical Strut Allograft Plates and Conventional Metal Plates

46

Citations

16

References

2003

Year

Abstract

This study compared the stability of periprosthetic femur fractures fixed using cortical allograft struts with a metal plate. Cadaveric specimens were loaded in single-leg stance and stair climbiing to 2250 N. Optimum stability in single-leg stance was achieved with two long struts medially and laterally. No clear advantage was noted in using a second strut in stair climbing. Cables rather than wires were useful in single-leg stance, but not in stair climbing. Allograft cortical struts are a biomechanically sound alternative to metal plates fixed with screws and cables for femur fracture fixation below a well-fixed femoral component.

References

YearCitations

Page 1