MedicineBiomedical EngineeringOncologyBiomedical OpticsBiophotonicsHistopathologyBiomedical Laser ApplicationsFibrosisHepatic FibrosisHepatic NecrosisLaser SurgeryLaser-tissue InteractionFibrotic ChangeFocal Laser PhotocoagulationLaser CoagulationLaser PhotocoagulationNd:yag Laser SurgeryNd:yag Laser TreatmentTissue VaporizationVaporization
Thirty-six deep hepatic lesions of localized photocoagulation were induced in 11 pigs by means of a neodymium-YAG laser. Laser applications of 80 W/10 sec (10.190 W/cm2) were transmitted through a handpiece coupled to a water-cooling circulation system to protect the quartz fiber and positioned through an echo-guided trocar. During irradiation, temperature was sufficient for vaporization up to 5 mm from the laser source and high enough for tumor cell kill at a 10-mm distance (54 degrees C/60 sec). Intraoperative ultrasound visualized increasing photocoagulation (12-18 mm), and further controls demonstrated an echo-free core of vaporization progressively covered by increasing fibrosis, well demarcated from normal parenchyma. Microscopy revealed central coagulative necrosis marginated from the third day by a growing fibrosis. By day 20 immunoblasts and mast cells were in profusion in the lesion border, and by day 120 a fibrotic network had invaded the scar and confirmed healing free of complication. This technique is proposed for deep vaporization of disseminated hepatic metastases.