Selective Dysfunction of Ventricular Electrode‐Endocardial Junction Following DC Cardioversion in a Patient with a Dual Chamber Pacemaker

Gopal D. Das, DALE B. STAFFANSON

Pacing and Clinical Electrophysiology · 1997 · 26 citations · 0 references

Concepts

Abstract

A patient with a dual chamber pacemaker received two transthoracic (anterior-anterior paddle position) shocks for VF. Subsequent analysis of stimulation thresholds revealed a marked rise in the ventricular threshold only. Later, she received three additional shocks (anterior-posterior paddle position) during electrophysiological study to terminate episodes of induced VT. Following these, the ventricular stimulation threshold was > 10 V at 2.0 ms. The selective damage at the ventricular site in this patient led us to carry out in vitro studies that revealed a preferential shunting of high electrical energy into the ventricular lead, as compared to the atrial lead, following a DC shock. These observations may explain the selective pacing malfunction observed in out patient.