Abstract

We show that the position of a fluorescent nanoparticle can be measured in three dimensions with subnanometer precision and 100-ms temporal resolution by use of standard epifluorescence video imaging in off-focus mode. The particle can be tracked without feedback in a volume of at least 40 microm x 60 microm x 3 microm. With the technique presented, the structure-mobility relationship of 216-nm latex particle in a porous polymer network was studied in three dimensions.