Abstract

Retinal thickness measurements in subjects with DME obtained using Spectralis OCT are considerably less variable than has been reported with other devices. Changes in central subfield thickness >8 μm can be considered more indicative of true clinical change rather than measurement variability. This finding informs clinical practice and clinical trial design.