Publication | Closed Access
A 1V 22µW 32-channel implantable EEG recording IC
30
Citations
7
References
2010
Year
Unknown Venue
Medical ElectronicsEngineeringAnalog DesignIntegrated CircuitsWireless Implantable DeviceMedical InstrumentationMixed-signal Integrated CircuitNoiseW 32-ChannelAdjustable Gain StagesAnalog-to-digital ConverterElectrical EngineeringData ConverterB Sar AdcComputer EngineeringMicroelectronicsNeural InterfaceBrain-computer InterfaceBioelectronicsElectrophysiologyImplantable Eeg
An implantable EEG recording IC including 32 analog low noise amplifiers, band-pass filters, adjustable gain stages, and a 10 b SAR ADC is presented. The chip, implemented in 0.35 μm CMOS process, achieves 1.15 μV <sub xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">rms</sub> total noise in a 0.5-to-150 Hz band and consumes 22 μW from a 1 V supply, attaining a noise efficiency factor of 2.24.