Assessment of the setup dependence of detector response functions for mega‐voltage linear accelerators

26

Citations

15

References

2010

Year

Abstract

The authors have shown that the detector response varies with detector type, depth, SSD, and detector scan direction. In-plane vs. cross-plane scanning can require calculation of a direction dependent response function. The effect of a 10% overall variation in the response function, for an ionization chamber, translates to a small deviation in the penumbra from that of the Edge detector measured profile when deconvolved. Due to the uncertainties introduced by deconvolution the Edge detector would be preferable in obtaining an approximation of the true profile, particularly for field sizes where the energy dependence of the diode can be neglected. However, an averaged response function could be utilized to provide a good approximation of the true profile for large ionization chambers and for larger fields for which diode detectors are not recommended.