Liquid chromatography with electrospray ion‐trap mass spectrometry for the determination of anatoxins in cyanobacteria and drinking water

Ambrose Furey, Janet Crowley, Mary Lehane, Kevin J. James

Rapid Communications in Mass Spectrometry · 2003 · 70 citations · 29 references

Concepts

Abstract

Anatoxin-a (AN) and homoanatoxin-a (HMAN) are potent neurotoxins produced by a number of cyanobacterial species. A new, sensitive liquid chromatography/multiple tandem mass spectrometry (LC/MS(n)) method has been developed for the determination of these neurotoxins. The LC system was coupled, via an electrospray ionisation (ESI) source, to an ion-trap mass spectrometer in positive ion mode. The [M+H](+) ions at m/z 166 (anatoxin-a) and m/z 180 (homoanatoxin-a) were used as the precursor ions for multiple MS experiments. MS(2)bond;MS(4) spectra displayed major fragment ions at m/z 149 (AN), 163 (HMAN), assigned to [Mbond;NH(3)+H](+); m/z 131 (AN), 145 (HMAN), assigned to [Mbond;NH(3)bond;H(2)O+H](+), and m/z 91 [C(7)H(7)](+). Although the chromatographic separation of these neurotoxins is problematic, reversed-phase LC, using a C(18) Luna column, proved successful. Calibration data for anatoxin-a using spiked water samples (10 mL) in LC/MS(n) modes were: LC/MS (25-1000 microg/L), r(2) = 0.998; LC/MS(2) (5-1000(microg/L), r(2) = 0.9993; LC/MS(3) (2.5-1000 microg/L), r(2) = 0.9997. Reproducibility data (% RSD, N = 3) for each LC/MS(n) mode ranged between 2.0 at 500 microg/L and 7.0 at 10 microg/L. The detection limit (S/N = 3) for AN was better than 0.03 ng (on-column) for LC/MS(3) which corresponded to 0.6 microg/L.