Metrologia · 2014 · 75 citations · 29 references
Natural SciencesPhysicsElectromagnetismInstrumentationNuclear InstrumentationSeismic InstrumentationVibration MeasurementZeeman EffectAbsolute GravimetersAccuracyActive Vibration IsolationMagnetic ShieldingQuadratic Zeeman EffectZeeman Effect MeasurementsZeeman Shift
Typical atomic gravimeters have a heavy magnetic shielding to avoid the quadratic Zeeman effect, and also a complicated active vibration isolation system to suppress the vibration noise. Both of them make it difficult to build a mobile and compact gravimeter. In this paper, we present the implementation of an atomic gravimeter aiming at field applications. Our gravimeter uses improved magnetic coils instead of the expensive mu-metal for magnetic shielding. The quadratic Zeeman shift is evaluated with high accuracy. Moreover, a portable platform of relatively small size is applied for vibration isolation. The total interrogation time is optimized to 120 ms and the repetition rate is 2.2 Hz. A sensitivity of 1.0 × 10−7 g Hz−1/2 and a resolution of 5.7 × 10−9 g within 1000 s integration time are reached. A continuous g measurement over 128 h is carried out. Moreover, a whole seismic wave of about 1 h that occurred in Pakistan on 28 September 2013 is recorded by our atomic gravimeter. The results coincide with that recorded by a traditional seismic detector very well.
29
Atomic interferometry using stimulated Raman transitions
Mark A. Kasevich, Steven Chu · Physical Review Letters · 1991 · 1.3K citations
High-precision gravity measurements using atom interferometry
Achim Peters, Keng Yeow Chung, Steven Chu · Metrologia · 2001 · 939 citations
Sensitive absolute-gravity gradiometry using atom interferometry
J. M. McGuirk, G. T. Foster, J. B. Fixler et al. · Physical Review A · 2002 · 571 citations · Full text
Atom Interferometer Measurement of the Newtonian Constant of Gravity
J. B. Fixler, G. T. Foster, J. M. McGuirk et al. · Science · 2007 · 542 citations