Calibration of helical tomotherapy machine using EPR/alanine dosimetry
14
Citations
18
References
2011
Year
Beam dose rate estimation results were found to be in good agreement with the reference value given by the manufacturer at 2% standard uncertainty. Moreover, the dose determination method was set up with a deviation around 2% (at a 2% standard uncertainty).