A New No-Lyse, No-Wash Flow-Cytometric Method for the Determination of CD4 T Cells in Blood Samples

Burkhard Greve, U. Cassens, Christine Möller Westerberg, Walter Sibrowski, D. Reichelt, W. Göhde

Transfusion Medicine and Hemotherapy · 2003 · 37 citations · 11 references

Concepts

Abstract

<b>Background: </b>Commonly used flow-cytometric methods for immunophenotyping are based on erythrocyte lysing reagents. It is known from the literature that these reagents result in a significant loss of leucocytes caused by membrane destruction. Although the dual-platform method should compensate this phenomenon, the subset-specific individual differences in sensitivity to lysing reagents lead to incorrect values. In order to overcome this problem we introduce a no-lyse, no-wash procedure in combination with absolute true volumetric counting (TVC) for the enumeration of CD4 T cells. <b>Material and Methods: </b>Whole-blood samples of 50 blood donors and 20 samples of patients with acquired immunodeficiency syndrome (AIDS) were treated with both a lyse, no-wash and a nolyse, no-wash procedure. Then, CD4 T cells were counted with a TVC flow cytometer (CyFlow Counter). 30 blood samples of blood donors were treated with a lyse and wash protocol and measured by the CyFlow Counter and FACS-Calibur system (reference method). A new gating strategy was used for the data analysis for both, the lyse and no-lyse method and compared to the traditional gating procedure. <b>Results: </b>The TVC method showed a good reproducibility for both, the lyse (CV 1.9%) and the no-lyse procedure (CV 1.5%). CD4 counts measured by the no-lyse procedure are on average 10% higher than by using the lysing protocol. The comparison of the CyFlow Counter and FACS-Calibur results showed a correlation of r = 0.961. The simplified gating strategy shows a good correlation to the traditional gating procedure (r = 0.998). <b>Conclusion: </b>Erythrocyte lysing procedures cause substantial cell loss with individual values for every single subclass and patient. Therefore, the use of a no-lyse procedure is recommended. The new gating strategy is fully comparable to the traditional one and simplifies enumeration of CD4 T cells. Using the no-lyse procedure in combination with the new gating strategy, it is possible to reduce the costs per sample from EUR 30.– to below EUR 2.–.

References

11