Comparative Validation of Quantitative Coronary Angiography Systems
169
Citations
46
References
1995
Year
There is a marked variability in performance between systems when assessed over the range of 0.5 to 1.9 mm. The range of accuracy, intercept, and slope values of this report indicates that absolute measurements of luminal diameter from different multicenter angiographic trials may not be directly comparable and additionally suggests that such absolute measurements may not be directly applicable to clinical practice using an on-line QCA system with a different edge detection algorithm. Power calculations and study design of angiographic trials should be adjusted for the precision of the QCA system used to avoid the risk of failing to detect small differences in patient populations. This study may guide the fine-tuning of algorithms incorporated within each system and facilitate the maintenance of high standards of QCA for scientific studies.